1. Under the Exam Window. Select Close Study Icon 2. Select Close Study if you are done taking X-rays Select Suspend if you need to take more X-rays to that Patient.
Replace Battery 1. To replace battery please firstly remove footers and unscrew the rear panel. 2. Then unscrew the battery holder and unplug tiny wires. We recommend to take pictures before to be sure of proper connecting later. The wires in the red ...
EV PANELS ONLY 1. Panel Connection - Cannot Connect to the Detector Message. 2. Detector ICON shows disconnected 3. NOTE READY light shows RED. 4. On Wireless Panel- Make sure the Panel is Turnned ON and the BLUE LIGHT is flashing Make sure the ...
1. Dicom error. 2. Clear Dicom error and Re-send Image Select Red Arrow Select AbortALL 3. resend image if error continues go to Step 4. 4. Verify the Dicom destination is correct. 5. Select Setting-PACS-PACSServer-Echo If you get an echo error means ...
How to recover lost studies from the "StudyList". How to recover lost images in the “ExamVue Duo” StudyList? 1. If some of the exams you have been working on in the ExamVue Duo software still do not appear in the “StudyList”, it is very possible that ...